166
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.8 Network Errors
Code Error name Countermeasures, etc.
20000 Memory error Open and close the door, and then con- rm operation. If re-occurrence
seems likely, replace the SCB2 and CF card.
21000 Hardware error Open and close the door, and then con- rm operation. If re-occurrence
seems likely, replace the SCB2 and CF card.
26000 Invalid Job Error A problem such as absence of a search job has occurred and execution
is not possible.
26100 Mode-out Job Error The search job is invalid for Standby, Read, and Maintenance modes,
and job execution is not possible.
26200 Invalid Console Check the Console is correctly operating.
One or more of the search job's registration or return address consoles
is invalid, and job execution is not possible
26300 Erasure Job Error Check the same job can be executed after erasure.
During erasure by the mechanism section, job execution is not possible.