155
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.3 Mechanical Control Error Code (04101 ~ 04106)
04101 ~ 04106 indicates an error occurred on the motors.
Error
code
Error name Error description Action
04101
Subscan motor NG A deviation occurred between the
elevation pulse number and the
descent pulse number (i.e., HP
motion) during initialization opera-
tions for subscanning.
Check MCB2:CN5-, pins 16 through 18
(Subscan HP Sensor) and MCB2:CN6
(Subscan Motor)
04102
Transportation Motor
NG
A deviation occurred between the
pulse number for [insertion posi-
tion -->Escape Position motion]
and the pulse number for [Escape
Position --> insertion position
motion] during initialization opera-
tions for the Transport Unit.
Check MCB2:CN11-, pins 32 through 34
(Feed HP sensor); MCB2:CN11-, pins
35 through 37 (Feed excape position
sensor); and MCB2:CN15 (Transporta-
tion Motor)
04103
Elevator motor NG A deviation occurred between the
descent pulse number and the
elevation pulse number (i.e., HP
motion) during initialization opera-
tions for the elevation bed.
Check MCB2:CN11-, pins 26 through 28
(Elevator HP Sensor) and MCB2:CN16
(Elevator Motor)
04104
Justication motor NG A deviation occurred between the
closing-direction pulse number
and the opening-direction pulse
number (i.e., HP motion) during
initialization operations for justi-
cation.
Check MCB2:CN11-, pins 29 through 31
(Justication HP Sensor) and
MCB2:CN17 (Justication Motor)
04105
Insertion shutter open/
close time-out
A time out occurred due to lack of
response from the sensor upon
opening or closing of the installa-
tion port shutters.
Check MCB2:CN10-, pins 11 through 13
(Insertion Shutter Open Sensor);
MCB2:CN10-, pins 14 through 16 (shut-
ter closed sensor); and MCB2:CN10-,
pins 3 through 4 (Shutter Motor)
04106 Lock time-out A time out occurred due to lack of
response from the sensor upon
locking.
Check MCB2:CN11-, pins 3 through
5 (Lock Release Sensor) and
MCB2:CN11-, pins 1 and 2 (Lock
Release Motor)
Check the plunger pin is actuating. if
it is not smoothly action
ig, replace
the cassette receive unit. (p.54)

Table of Contents

Other manuals for Konica Minolta REGIUS 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 190 and is the answer not in the manual?

Konica Minolta REGIUS 190 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals