4.3 Troubleshooting for Image Quality
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.3.4 Case 3: Horizontal Streaks in Image
Phenomena Cause/Remedy
1 Horizontal streak occurring at one location
within the image. *
There is a possibility that an external signal is being
applied during reading.
Ensure that the device is protected from external inu-
ences during reading, and perform exposure again.
2
Horizontal lines extend from shielding
materials such as lead.
This can happen when the lead or shielding material is
in a blank section.
This phenomenon happens only under very specic
conditions, and it can be resolved by moving the shield-
ing material as far as possible from the subject or by
restricting the exposure eld.
If this continues to occur repeatedly, please contact the
plant or Development.
3 Sudden change in brightness approximately
30 mm from the bottom of the image.
Return the plate with warping or peeling to the REGIUS
Console.
If the problem repeats with different plates, adjust the
peel detector roller.(p.117)
Plate warping or peeling can lead to contact between
the plate and the detection roller.
4 Regular, diagonal stripes This can occur as a
result of light leaking from the exterior.
Light may be leaking in the device.
Reassemble the exterior covering or darken the room,
and then take a solid image to conrm quality

Table of Contents

Other manuals for Konica Minolta REGIUS 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 190 and is the answer not in the manual?

Konica Minolta REGIUS 190 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals