20
1.5 Description of Operation
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
1.5.2 Normal Operation
1 When a cassette is inserted, the light beams from the inser-
tion slot sensors' emission side is interrupted, and the inser-
tion slot sensors' receptor side turns on.
Two individual insertion slot sensors with emitters and
receptors are arranged in parallel, and when a small cas-
sette is inserted along right side, only one of these sensors
will turn on and an error will be generated.
2 When the insertion slot sensors turn on, the insertion motor
will operate for one second.
The cassette is pulled in by rollers, and reading by the bar
code reader commences.
An error will be generated if the bar code on the back plate
cannot be read.
3 The shutter motor starts to operate and the shutter at the
bottom of the insertion slot opens up.
The elevation stand moves up to a receiving position suit-
able for the size of the cassette.
The elevation stand moves down to the feed position and the
cassette is drawn further into the device.
Several seconds after the insertion slot sensors turn off, the
shutter closes.
When the cassette has been fully drawn into the transport
section, the width justication motor drives the width justi-
cation mechanism so that the cassette is held on the left and
right.
A cassette which was inserted aligned to the left will be
aligned to the center at this time.
4 The feed motor starts to operate and the transport section
moves to the absorb position.
The feed sensor turns on.
Transport section then moves to the absorb position.
Back plate is stuck to the Cassette Magnetizing Plate.
5 The transport section's lock release motor starts to operate
and the push-up pin pushes against the cassette's open/
close lock, raising it up.
The feed motor operates and the transport section is
returned to the retracted position.
The back plate and the REGIUS Plate become attached to
the absorption plate, with only the front plate remaining in
the transport section.
Insertion slot sensor
Cassette insertion
Insertion motor (M1)
Bar code reader
Cassette Receive Sensor (S2)
Elevator Motor (PM1)
Justification Motor (PM2)
Shutter Motor (M2)
Justification Sensor-L (S3)/
Justification Sensor-R (S4)
Feed Motor (PM3)
Feed Sensor (S11)
Lock-Release Motor (M4)
Front Plate
Back Plate & REGIUS Plate

Table of Contents

Other manuals for Konica Minolta REGIUS 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 190 and is the answer not in the manual?

Konica Minolta REGIUS 190 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals