150
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4 Error Code
4.4.1 General Procedures Required when Error Occurs
Below is the general procedure required when an error occurs.
Collect the log.
When an error is displayed on the LCD of REGIUS 190, collect the error log using REGIUS Console.
Reset
1.
Open / close the front cover.
2.
Remove the cassette if it is capable.
3.
Press "operation" button, and shut down the REGIUS 190.
4.
Tu rn OFF the circuit breaker located on the side of the main
unit.
5.
Restart the REGIUS 190.
Information to be attached when sending the error log
When sending the error log to the manufacturer, attach the related information listed below as detailed as possi-
ble.
Network Condition at the institute JM: External or internal?
n: m
Any dedicated reader is connected?
IP address /subnet mask
REGIUS 190
•JM
REGIUS Console
Reset Reset implemented by opening / closing the door.
Needed to turn the breaker OFF / ON.
Others, details

Table of Contents

Other manuals for Konica Minolta REGIUS 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 190 and is the answer not in the manual?

Konica Minolta REGIUS 190 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals