161
4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.6 Mechanical Control Error Code (04900 ~ 04908)
Type Error name Error description 04900~ 04908 indicates an error which can not be classified.
Error
code
Error name Error description Action
04900 Cassette exists when
initializing (Cassette
receive Unit)
The Cassette receive sensor was
On at the start of initialization
operations.
Open the front door, remove any cas-
settes, and then close the door and
restore. If restoration was not possible
(i.e., Ready is not achieved after com-
pletion of initialization operations), press
the Operation switch and perform a
restart after the power has turned off.
04901 Cassette exists when
initializing (justifica-
tion)
The justification sensor-L or -R
was On at the start of initialization
operations.
Open the front door, remove any cas-
settes, and then close the door and
restore. If restoration was not possible
(i.e., Ready is not achieved after com-
pletion of initialization operations), press
the Operation switch and perform a
restart after the power has turned off.
04902 Cassette exists when
initializing (ejection
shutter open)
The eject shutter close sensor
was Off at the start of initialization
operations.
Open the front door, remove any cas-
settes, and then close the door and
restore. If restoration was not possible
(i.e., Ready is not achieved after com-
pletion of initialization operations), press
the Operation switch and perform a
restart after the power has turned off.
04903 Cassette exists when
initializing (absorbed
on back plate)
The back plate's Absorption
Detect sensor was On at the start
of initialization operations.
Open the front door, remove any cas-
settes, and then close the door and
restore. If restoration was not possible
(i.e., Ready is not achieved after com-
pletion of initialization operations), press
the Operation switch and perform a
restart after the power has turned off.
04904 Erase lamp down (2
lamps)
Both of the erasure lamps have
blown out.
Carry out lamp replacement immedi-
ately.
04905 SCB2 communication
error
No ACK was received in
response to an ENQ code in
accordance with the SCB2 com-
munication protocol.
If the error repeats after restart by per-
forming open/close of the front cover,
check the follows.
1) Reoccurrence even after opening
and closing of the door and a restart.
2) Check the connection of the power-
supply connector and the SCB2.
3) Check the voltage level in the SCB2.
4) Confirm that the SCB2(CN5)-
MCB2(CN4) connectors are con-
nected.
5) Confirm that the CF card has been
inserted.
6) Replace the CF card and check the
subsequent operation.
7) Replace the SCB2 and confirm the
subsequent operation.