4.3 Troubleshooting for Image Quality
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.3.23 Case 22: Image duplication
4.3.24 Case 23: Sensitivity correction is not within the specified range
4.3.25 Case 24: All post-processing images have high contrast.
4.3.26 Case 25: All post-processing images have low contrast.
Remedy
1 Check the erase lamp is operating during the erasure process.
2 •Erase and then carry out reading without using any correction.
If the problem repeats, replace the pate (as large doses have caused the plate to break).
This operation is to be carried out with correction.
Remedy
1 Check whether the correction image's exposure conditions and input values are correct.
Phenomena Cause/Remedy
1 A high overall contrast is set for the process-
ing parameter of output density.
Set the output density "H" of G-Value/Shift parameters to
a lower value.
Set the output density "L" of G-Value/Shift parameters to
higher value.
For more details, refer to the Image Processing Adjust-
ment Manual.
Phenomena Cause/Remedy
1 A low overall contrast is set for the process-
ing parameter of output density.
Set the output density "H" of G-Value/Shift parameters to
a higher value.
Set the output density "L" of G-Value/Shift parameters to
a lower value.
For more details, refer to the Image Processing Adjust-
ment Manual.

Table of Contents

Other manuals for Konica Minolta REGIUS 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 190 and is the answer not in the manual?

Konica Minolta REGIUS 190 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals