4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.7 SCB2 Error Codes
Code Medium Classifi-
cation
Detailed Clas-
sification
Countermeasures, etc.
10000 Memory error SCB2 replacement or CF card
replacement
11000 Hardware error SCB2 replacement
12000 Internal communi-
cation error
Mechanical
control com-
munication
High voltage ON
Timeout
Conrm that there was no continuous
insertion.
• Open and close the door and then
restore. If restoration is not possi-
ble, restart.
12001 Internal error Mechanical
control com-
munication
Laser ON Timeout Open and close the door and then
restore. If restoration is not possible,
restart. Following restart, conrm the
following.
LD peripherals (LD power supply
and optical unit)
Check the wiring on the board.
12002 Internal error Mechanical
control com-
munication
Mech Ready Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart.
12003 Internal error Mechanical
control com-
munication
Standby Timeout Open and close the door and then
restore. If restoration is not possible,
restart.
12010 Internal error Mechanical
control com-
munication
Resent counts
Overow
Open and close the door and then
restore. If restoration is not possible,
restart.
12011 Internal error Mechanical
control com-
munication
CRC Erorr counts
Over-ow
Open and close the door and then
restore. If restoration is not possible,
restart.
12200 Internal error Network com-
munication
Net Initialize Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart. If the problem then reoccurs,
conrm whether the JM is operating.
Also conrm that the device is con-
nected to the net.
12201 Internal error Network com-
munication
Net Ready Timeout Open and close the door and then
restore. If restoration is not possible,
restart.
12202 Internal error Network com-
munication
Put Shading Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart.
12203 Internal error Network com-
munication
Status Check Time-
out
Open and close the door and then
restore. If restoration is not possible,
restart.
12204 Internal error Network com-
munication
Barcode Seek Tim-
eout
Open and close the door and then
restore. If restoration is not possible,
restart.

Table of Contents

Other manuals for Konica Minolta REGIUS 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta REGIUS 190 and is the answer not in the manual?

Konica Minolta REGIUS 190 Specifications

General IconGeneral
BrandKonica Minolta
ModelREGIUS 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals