3.9 Adjusting the Position of Insertion Guide Switch (SW4)
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.9 Adjusting the Position of Insertion Guide Switch (SW4)
If the assembled positions of the insertion guide SW and the
receiver plate of the magnet catch are incorrect, initializing will
not be initiated even the insertion guide is put back to the origi-
nal position.
When the insertion guide or the insertion guide switch (SW4) is
replaced, follow the procedures below to check and adjust the
assembled position.
1.
Loosen two fixing screws at right and left on the magnet
catch receiver plate.
2.
Lower the insertion guide (bossed part) beneath the slit.
3.
Tighten two fixing screws (right and left) to secure the mag-
net catch receiver plate.
4.
Pull up the insertion guide.
5.
Loosen 2 fixing screws which are securing the insertio
guide switch (SW4).
Impor
tant Loosen the screws as slightly as possi-
ble. Loosening too far will allow the nut
on the nut plate on the back to fall off.
6.
Fluctuate the reverse-actuating hinge lever, and measure
the distance "A" when the insertion guide switch (SW4) is
tuned on.
Criterion : When A = 13.5mm, SW4 should be turned on.
7.
Adjust the fixing screws (2pcs) to bring the position of
insertion guide switch (SW4) so that it falls complies with
the criterion.
Reverse-Actuating Hinge Lever
Fixing Screw (Rigth)
Slit
Insertion Guide Switch
(SW4)
Magnet Catch Receiver (Right)
Reverse-Actuating Hinge Lever
A
Insertion Guide Switch (SW4)