4.4 Error Code
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
4.4.4 Mechanical Control Error Code (04150 ~ 04165)
04150 ~ 04165 indicates an error occurred on the firmware.
Error
code
Error name Error description Action
04150
Lock initializing pro-
cess
Initialization of the lock motor
cannot be carried out as a result
of a problem in the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04151
Insertion shutter initial-
izing process
Initialization of the insertion slot
shutter motor cannot be carried
out as a result of a problem in the
firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04152
Cassette receive Unit
initializing process
Initialization of the elevation-bed
motor cannot be carried out as a
result of a problem in the firm-
ware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04153
Justification initializing
process
Initialization of the justification
motor cannot be carried out as a
result of a problem in the firm-
ware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04154
Transport Unit initializ-
ing process
Initialization of the Transportation
Motor cannot be carried out as a
result of a problem in the firm-
ware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04155
Subscan initializing
process
Initialization of the auxiliary scan-
ning motor cannot be carried out
as a result of a problem in the
firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04156
Feed in operation 1
start NG
Draw-in operation 1 could not be
carried out by the insertion slot
motor as a result of a problem in
the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04157
Feed in operation 2
start NG
Draw-in operation 2 could not be
carried out by the insertion slot
motor as a result of a problem in
the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04158
Insertion slot ejection
start NG
An eject operation could not be
carried out by the insertion slot
motor as a result of a problem in
the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04159
Ejection start NG An eject operation could not be
carried out by the removal-port
motor as a result of a problem in
the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04160
Insertion slot shutter
open NG
An opening operation could not
be carried out by the insertion slot
motor as a result of a problem in
the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.
04161
Insertion slot shutter
close NG
A closing operation could not be
carried out by the insertion slot
shutter motor as a result of a
problem in the firmware.
Open the front door, remove any cas-
sette that may be present inside the
equipment, and then carry out a restart.