124
3.8 Checking and Adjusting the Justification Allowance
REGIUS MODEL 190 Service Manual Ver.1.00 2004.11.01
3.8 Checking and Adjusting the Justication Allowance
Set the 14" x 17" cassette on the justication mechanism, and
measure the gap between the justication control guide-R1 and
justication control guide-R2.
Criterion: 1 ~ 3mm
If the drive pulley is loose, tighten the drive pulley at the
force of 0.6N•cm (6kgf•cm) using a torque driver.
Justification Control
Guide-R1
Left Right
Justification Control
Guide-R2
Gap
3 Different Aspects of
Justification Mechanism

Table of Contents