Page #306 background imageLoading...
Page #306 background image
Model 2450
Interactive SourceMeter® Instrument Reference Manual Section 4: Source-
measure considerations
2450-901-01 Rev. B/September 2013 4-27
Setting Default value on
reset
Not available from front panel
[:SENSe[1]]:<function>:RELative (on page 6-55)
smu.measure.rel.level (on page 8-122)
0
MENU > Measure > Filter/Math > Rel State
[:SENSe[1]]:<function>:RELative:STATe (on page 6-57)
smu.measure.rel.enable (on page 8-121)
Off
MENU > Measure > Settings > Sense Mode
[:SENSe[1]]:<function>:RSENse (on page 6-58)
smu.measure.sense (on page 8-123)
2-wire
Not available from front panel
[:SENSe[1]]:<function>:UNIT (on page 6-59)
smu.measure.unit (on page 8-125)
Current: Amp
Voltage: Voltage
Resistance: Ohms
Source reset values
Setting Default value on
reset
Not available from front panel
:SOURce[1]:<function>:DELay:AUTO (on page 6-67)
smu.source.autodelay (on page 8-128)
On
MENU > Source > Settings > Hi Capacitance
:SOURce[1]:<function>:HIGH:CAPacitance (on page 6-69)
smu.source.highc (on page 8-135)
Off
HOME > Source
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude] (on
page 6-70)
smu.source.level (on page 8-135)
0
HOME > Limit
:SOURce[1]:<function>:<x>LIMit[:LEVel] (on page 6-71)
smu.source.xlimit.level (on page 8-137)
Current: 105 µA
Voltage: 21 V
FUNCTION button
:SOURce[1]:FUNCtion[:MODE] (on page 6-72)
Not applicable for TSP
Current
MENU > Source > Settings > Overvoltage Protection
:SOURce[1]:<function>:PROTection[:LEVel] (on page 6-73)
smu.source.protect.level (on page 8-141)
None
MENU > Source > Settings > Source Range
:SOURce[1]:<function>:RANGe (on page 6-74)
smu.source.range (on page 8-143)
Current: 1e-08
Voltage: 20 mV
HOME > Source Range
:SOURce[1]:<function>:RANGe:AUTO (on page 6-75)
smu.source.autorange (on page 8-127)
On
MENU > Source > Settings > Source Readback
:SOURce[1]:<function>:READ:BACK (on page 6-77)
smu.source.readback (on page 8-144)
On
FUNCTION key
:SOURce[1]:FUNCtion[:MODE] (on page 6-72)
smu.source.func (on page 8-134)
Voltage

Table of Contents