:WGEN Commands 33
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 1015
:WGEN:MODulation:FM:DEViation
(see page 1138)
Command Syntax
:WGEN:MODulation:FM:DEViation <frequency>
<frequency> ::= frequency deviation in Hz in NR3 format
The :WGEN:MODulation:FM:DEViation command specifies the frequency
deviation from the original carrier signal frequency.
When the modulating signal is at its maximum amplitude, the output
frequency is the carrier signal frequency plus the deviation amount, and
when the modulating signal is at its minimum amplitude, the output
frequency is the carrier signal frequency minus the deviation amount.
The frequency deviation cannot be greater than the original carrier signal
frequency.
Also, the sum of the original carrier signal frequency and the frequency
deviation must be less than or equal to the maximum frequency for the
selected waveform generator function plus 100 kHz.
Query Syntax
:WGEN:MODulation:FM:DEViation?
The :WGEN:MODulation:FM:DEViation? query returns the frequency
deviation setting.
Return Format
<frequency><NL>
<frequency> ::= frequency deviation in Hz in NR3 format
See Also • ":WGEN:MODulation:AM:DEPTh" on page 1013
• ":WGEN:MODulation:AM:FREQuency" on page 1014
• ":WGEN:MODulation:FM:FREQuency" on page 1016
• ":WGEN:MODulation:FSKey:FREQuency" on page 1017
• ":WGEN:MODulation:FSKey:RATE" on page 1018
• ":WGEN:MODulation:FUNCtion" on page 1019
• ":WGEN:MODulation:FUNCtion:RAMP:SYMMetry" on page 1020
• ":WGEN:MODulation:STATe" on page 1022
• ":WGEN:MODulation:TYPE" on page 1023