556 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :POWer Commands
:POWer:ONOFf:TEST
(see page 1138)
Command Syntax
:POWer:ONOFf:TEST {{0 | OFF} | {1 | ON}}
The :POWer:ONOFf:TEST command selects whether turn on or turn off
analysis is performed:
• ON — Turn On — measures the time taken to get the output voltage of
the power supply after the input voltage is applied.
• OFF — Turn Off — measures the time taken for the output voltage of
the power supply to turn off after the input voltage is removed.
Query Syntax
:POWer:ONOFf:TEST?
The :POWer:ONOFf:TEST query returns the selected test type.
Return Format
{0 | 1}
See Also • ":POWer:ONOFf:APPLy" on page 553
• ":POWer:ONOFf:EXIT" on page 554
• ":POWer:ONOFf:NEXT" on page 555

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals