926 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
31 :TRIGger Commands
:TRIGger:PATTern:RANGe
(see page 1138)
Command Syntax
:TRIGger:PATTern:RANGe <less_than_time>[<suffix>],
<greater_than_time>[<suffix>]
<greater_than_time> ::= 10 ns to 9.99 seconds in NR3 format
<less_than_time> ::= 15 ns to 10 seconds in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:PATTern:RANGe command sets the duration for the defined
pattern when the :TRIGger:PATTern:QUALifier command is set to INRange
or OUTRange. You can enter the parameters in any order — the smaller
value becomes the <greater_than_time> and the larger value becomes the
<less_than_time>.
Query Syntax
:TRIGger:PATTern:RANGe?
The :TRIGger:PATTern:RANGe? query returns the duration time for the
defined pattern.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also • "Introduction to :TRIGger Commands" on page 877
• ":TRIGger:PATTern" on page 919
• ":TRIGger:PATTern:QUALifier" on page 924
• ":TRIGger:MODE" on page 886