Getting Started 3
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 71
Other Ways of Sending Commands
Standard Commands for Programmable Instrumentation (SCPI) can also be
sent via a Telnet socket or through the Browser Web Control:
• "Telnet Sockets" on page 71
• "Sending SCPI Commands Using Browser Web Control" on page 71
Telnet Sockets
The following information is provided for programmers who wish to
control the oscilloscope with SCPI commands in a Telnet session.
To connect to the oscilloscope via a telnet socket, issue the following
command:
telnet <hostname> 5024
where <hostname> is the hostname of the oscilloscope. This will give you a
command line with prompt.
For a command line without a prompt, use port 5025. For example:
telnet <hostname> 5025
Sending SCPI Commands Using Browser Web Control
To send SCPI commands using the Browser Web Control feature, establish
a connection to the oscilloscope via LAN as described in the InfiniiVision
3000 X- Series Oscilloscopes User's Guide. When you make the connection
to the oscilloscope via LAN and the instrument's welcome page is
displayed, select the Browser Web Control tab, then select the Remote
Programming link.