:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 685
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition
(see page 1138)
Command Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition <cycle_count_repetition>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer values 2, 4, 8, 16, 32, or 64
The :SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition command sets the
repetition number of the FlexRay cycle count (in the frame header) to
trigger on. The cycle count repetition setting is only valid when the
FlexRay trigger mode is set to FRAME.
Query Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition?
The :SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition? query returns the
current cycle count repetition setting for the FlexRay frame trigger setup.
Return Format
<cycle_count_repetition><NL>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer values 2, 4, 8, 16, 32, or 64
See Also • "Introduction to :TRIGger Commands" on page 877
• ":SBUS<n>:FLEXray:TRIGger" on page 679

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals