:TIMebase Commands 30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 875
:TIMebase:WINDow:SCALe
(see page 1138)
Command Syntax
:TIMebase:WINDow:SCALe <scale_value>
<scale_value> ::= scale value in seconds in NR3 format
The :TIMebase:WINDow:SCALe command sets the zoomed (delayed)
window horizontal scale (seconds/division). The main sweep scale
determines the range for this command. The maximum value is one half of
the :TIMebase:SCALe value.
Query Syntax
:TIMebase:WINDow:SCALe?
The :TIMebase:WINDow:SCALe? query returns the current zoomed window
scale setting.
Return Format
<scale_value><NL>
<scale_value> ::= current seconds per division for the zoomed window
See Also • "Introduction to :TIMebase Commands" on page 866
• ":TIMebase:RANGe" on page 869
• ":TIMebase:POSition" on page 868
• ":TIMebase:SCALe" on page 871
• ":TIMebase:WINDow:RANGe" on page 874