1016 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
33 :WGEN Commands
:WGEN:MODulation:FM:FREQuency
(see page 1138)
Command Syntax
:WGEN:MODulation:FM:FREQuency <frequency>
<frequency> ::= modulating waveform frequency in Hz in NR3 format
The :WGEN:MODulation:FM:FREQuency command specifies the frequency
of the modulating signal.
Query Syntax
:WGEN:MODulation:FM:FREQuency?
The :WGEN:MODulation:FM:FREQuency? query returns the frequency of the
modulating signal.
Return Format
<frequency><NL>
<frequency> ::= modulating waveform frequency in Hz in NR3 format
See Also • ":WGEN:MODulation:AM:DEPTh" on page 1013
• ":WGEN:MODulation:AM:FREQuency" on page 1014
• ":WGEN:MODulation:FM:DEViation" on page 1015
• ":WGEN:MODulation:FSKey:FREQuency" on page 1017
• ":WGEN:MODulation:FSKey:RATE" on page 1018
• ":WGEN:MODulation:FUNCtion" on page 1019
• ":WGEN:MODulation:FUNCtion:RAMP:SYMMetry" on page 1020
• ":WGEN:MODulation:STATe" on page 1022
• ":WGEN:MODulation:TYPE" on page 1023