1066 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:MEASure:LOWer
(see page 1138)
Command Syntax
:MEASure:LOWer <voltage>
The :MEASure:LOWer command sets the lower measurement threshold
value. This value and the UPPer value represent absolute values when the
thresholds are ABSolute and percentage when the thresholds are PERCent
as defined by the :MEASure:DEFine THResholds command.
Query Syntax
:MEASure:LOWer?
The :MEASure:LOWer? query returns the current lower threshold level.
Return Format
<voltage><NL>
<voltage> ::= the user-defined lower threshold in volts in NR3 format
See Also • "Introduction to :MEASure Commands" on page 406
• ":MEASure:THResholds" on page 1069
• ":MEASure:UPPer" on page 1076
NOTE
The :MEASure:LOWer command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 413)
instead.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals