1020 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
33 :WGEN Commands
:WGEN:MODulation:FUNCtion:RAMP:SYMMetry
(see page 1138)
Command Syntax
:WGEN:MODulation:FUNCtion:RAMP:SYMMetry <percent>
<percent> ::= symmetry percentage from 0% to 100% in NR1 format
The :WGEN:MODulation:FUNCtion:RAMP:SYMMetry command specifies the
amount of time per cycle that the ramp waveform is rising. The ramp
modulating waveform shape is specified with the
:WGEN:MODulation:FUNCtion command.
Query Syntax
:WGEN:MODulation:FUNCtion:RAMP:SYMMetry?
The :WGEN:MODulation:FUNCtion:RAMP:SYMMetry? query returns ramp
symmetry percentage setting.
Return Format
<percent><NL>
<percent> ::= symmetry percentage from 0% to 100% in NR1 format
See Also • ":WGEN:MODulation:AM:DEPTh" on page 1013
• ":WGEN:MODulation:AM:FREQuency" on page 1014
• ":WGEN:MODulation:FM:DEViation" on page 1015
• ":WGEN:MODulation:FM:FREQuency" on page 1016
• ":WGEN:MODulation:FSKey:FREQuency" on page 1017
• ":WGEN:MODulation:FSKey:RATE" on page 1018
• ":WGEN:MODulation:FUNCtion" on page 1019
• ":WGEN:MODulation:STATe" on page 1022
• ":WGEN:MODulation:TYPE" on page 1023