EasyManuals Logo
Home>Agilent Technologies>Test Equipment>InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series Programmer's Guide

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1032 background imageLoading...
Page #1032 background image
1032 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
33 :WGEN Commands
:WGEN:VOLTage:OFFSet
(see page 1138)
Command Syntax
:WGEN:VOLTage:OFFSet <offset>
<offset> ::= offset in volts in NR3 format
The :WGEN:VOLTage:OFFSet command specifies the waveform's offset
voltage or the DC level. Use the :WGEN:VOLTage command to specify the
amplitude.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example,
an amplitude of 5 V and an offset of 1 V is the same as a high- level
voltage of 4 V and a low-level voltage of - 1 V.
Query Syntax
:WGEN:VOLTage:OFFSet?
The :WGEN:VOLTage:OFFSet? query returns the currently specified
waveform offset voltage.
Return Format
<offset><NL>
<offset> ::= offset in volts in NR3 format
See Also • "Introduction to :WGEN Commands" on page 998
• ":WGEN:FUNCtion" on page 1007
• ":WGEN:VOLTage" on page 1029
• ":WGEN:VOLTage:HIGH" on page 1030
• ":WGEN:VOLTage:LOW" on page 1031

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals