872 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
30 :TIMebase Commands
:TIMebase:VERNier
(see page 1138)
Command Syntax
:TIMebase:VERNier <vernier value>
<vernier value> ::= {{1 | ON} | {0 | OFF}
The :TIMebase:VERNier command specifies whether the time base control's
vernier (fine horizontal adjustment) setting is ON (1) or OFF (0).
Query Syntax
:TIMebase:VERNier?
The :TIMebase:VERNier? query returns the current state of the time base
control's vernier setting.
Return Format
<vernier value><NL>
<vernier value> ::= {0 | 1}