30 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 1138)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full- scale horizontal time
in seconds for the zoomed (delayed) window. The range is 10 times the
current zoomed view window seconds per division setting. The main sweep
range determines the range for this command. The maximum value is one
half of the :TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window
timebase range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also • "Introduction to :TIMebase Commands" on page 866
• ":TIMebase:RANGe" on page 869
• ":TIMebase:POSition" on page 868
• ":TIMebase:SCALe" on page 871