550 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :POWer Commands
:POWer:MODulation:APPLy
(see page 1138)
Command Syntax
:POWer:MODulation:APPLy
The :POWer:MODulation:APPLy command applies the selected modulation
analysis type (:POWer:MODulation:TYPE).
The Modulation analysis measures the control pulse signal to a switching
device (MOSFET) and observes the trending of the pulse width, duty cycle,
period, frequency, etc. of the control pulse signal.
See Also • ":POWer:MODulation:SOURce" on page 551
• ":POWer:MODulation:TYPE" on page 552
• ":MEASure:VAVerage" on page 457
• ":MEASure:VRMS" on page 463
• ":MEASure:VRATio" on page 462
• ":MEASure:PERiod" on page 433
• ":MEASure:FREQuency" on page 426
• ":MEASure:PWIDth" on page 437
• ":MEASure:NWIDth" on page 429
• ":MEASure:DUTYcycle" on page 424
• ":MEASure:RISetime" on page 441
• ":MEASure:FALLtime" on page 425

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals