1084 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:MTESt:RUMode
(see page 1138)
Command Syntax
:MTESt:RUMode {FORever | TIME,<seconds> | {WAVeforms,<wfm_count>}}
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
The :MTESt:RUMode command determines the termination conditions for
the mask test. The choices are FORever, TIME, or WAVeforms.
• FORever — runs the Mask Test until the test is turned off.
• TIME — sets the amount of time in seconds that a mask test will run
before it terminates. The <seconds> parameter is a real number from 1
to 86400 seconds.
• WAVeforms — sets the maximum number of waveforms that are required
before the mask test terminates. The <wfm_count> parameter indicates
the number of waveforms that are to be acquired; it is an integer from
1 to 1,000,000,000.
Query Syntax
:MTESt:RUMode?
The :MTESt:RUMode? query returns the currently selected termination
condition and value.
Return Format
{FOR | TIME,<seconds> | {WAV,<wfm_count>}}<NL>
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
See Also • "Introduction to :MTESt Commands" on page 491
• ":MTESt:RUMode:SOFailure" on page 1085
The :MTESt:RUMode command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MTESt:RMODe command (see page 508) instead.