:MEASure Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 419
:MEASure:DUAL:VAMPlitude
(see page 1138)
Overview This measurement is available with the N2820A high sensitivity current
probe when both the Primary and Secondary probe cables are used. This
measurement joins the Zoom In waveform data below the probe's clamp
level with Zoom Out waveform data above the probe's clamp level to create
the waveform on which the measurement is made.
Command Syntax
:MEASure:DUAL:VAMPlitude [<source1>][,<source2>]
<source1>,<source2> ::= CHANnel<n> with N2820A probe connected
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:DUAL:VAMPlitude command installs a screen measurement
and starts a vertical amplitude measurement.
If the optional source parameter(s) are specified, the currently specified
source(s) are modified.
Query Syntax
:MEASure:DUAL:VAMPlitude? [<source1>][,<source2>]
The :MEASure:DUAL:VAMPlitude? query measures and returns the vertical
amplitude of the waveform. To determine the amplitude, the instrument
measures Vtop and Vbase, then calculates the amplitude as follows:
vertical amplitude = Vtop - Vbase
Return Format
<value><NL>
<value> ::= the amplitude of the selected waveform in NR3 format
See Also • ":MEASure:DUAL:CHARge" on page 418
• ":MEASure:DUAL:VAVerage" on page 420
• ":MEASure:DUAL:VBASe" on page 421
• ":MEASure:DUAL:VPP" on page 422
• ":MEASure:DUAL:VRMS" on page 423
• "Introduction to :MEASure Commands" on page 406
• ":MEASure:SOURce" on page 444
• ":MEASure:VTOP" on page 465