:SAVE Commands 26
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 621
:SAVE:WAVeform[:STARt]
(see page 1138)
Command Syntax
:SAVE:WAVeform[:STARt] [<file_name>]
<file_name> ::= quoted ASCII string
The :SAVE:WAVeform[:STARt] command saves oscilloscope waveform data
to a file.
See Also • "Introduction to :SAVE Commands" on page 607
• ":SAVE:WAVeform:FORMat" on page 622
• ":SAVE:WAVeform:LENGth" on page 623
• ":SAVE:FILename" on page 609
• ":RECall:SETup[:STARt]" on page 603
NOTE
Be sure to set the :SAVE:WAVeform:FORMat before saving waveform data. If the format is
NONE, the save waveform command will not succeed.
NOTE
If a file extension is provided as part of a specified <file_name>, and it does not match the
extension expected by the format specified in :SAVE:WAVeform:FORMat, the format will be
changed if the extension is a valid waveform file extension.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals