500 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
22 :MTESt Commands
:MTESt:COUNt:FWAVeforms
(see page 1138)
Query Syntax
:MTESt:COUNt:FWAVeforms? [CHANnel<n>]
<n> ::= 1 to (# analog channels) in NR1 format
The :MTESt:COUNt:FWAVeforms? query returns the total number of failed
waveforms in the current mask test run. This count is for all regions and
all waveforms collected on the channel specified by the optional parameter
or collected on the currently specified source channel (:MTESt:SOURce) if
there is no parameter.
Return Format
<failed><NL>
<failed> ::= number of failed waveforms in NR1 format.
See Also • "Introduction to :MTESt Commands" on page 491
• ":MTESt:COUNt:WAVeforms" on page 503
• ":MTESt:COUNt:TIME" on page 502
• ":MTESt:COUNt:RESet" on page 501
• ":MTESt:SOURce" on page 521
Example Code • "Example Code" on page 491

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals