1072 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
35 Obsolete and Discontinued Commands
:MEASure:TSTArt
(see page 1138)
Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1
cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also • "Introduction to :MARKer Commands" on page 380
• "Introduction to :MEASure Commands" on page 406
• ":MARKer:X1Position" on page 382
• ":MARKer:X2Position" on page 384
• ":MARKer:XDELta" on page 386
• ":MEASure:TDELta" on page 1068
• ":MEASure:TSTOp" on page 1073
The short form of this command, TSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 1140). The normal short form "TST" would be the same for
both TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X1Position command (see page 382) instead.