:SEARch Commands 28
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 825
:SEARch:SERial:I2S:RANGe
(see page 1138)
Command Syntax
:SEARch:SERial:I2S:RANGe <lower>, <upper>
<lower> ::= 32-bit integer in signed decimal, <nondecimal>, or <string>
<upper> ::= 32-bit integer in signed decimal, <nondecimal>, or <string>
<nondecimal> ::= #Hnn...n where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn...n" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :SEARch:SERial:I2S:RANGe command specifies the data value range
when searching for I2S events in the INRange and OUTRange search
modes (set by the :SEARch:SERial:I2S:MODE command).
You can enter the parameters in any order — the smaller value becomes
the <lower> and the larger value becomes the <upper>.
Query Syntax
:SEARch:SERial:I2S:RANGe?
The :SEARch:SERial:I2S:RANGe? query returns the current data value
range setting.
Return Format
<lower>, <upper><NL>
<lower> ::= 32-bit integer in signed decimal
<upper> ::= 32-bit integer in signed decimal
See Also • Chapter 28, “:SEARch Commands,” starting on page 777
• ":SEARch:SERial:I2S:MODE" on page 822