480 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
21 :MEASure Power Commands
:MEASure:OFFTime
(see page 1138)
Command Syntax
:MEASure:OFFTime [<source1>][,<source2>]
<source1>, <source2> ::= {CHANnel<n>}
<n> ::= 1 to (# analog channels) in NR1 format
The :MEASure:OFFTime command installs an "off time" measurement on
screen.
Turn off time measures the difference of time between when the input AC
Voltage last falls to 10% of its maximum amplitude to the time when the
output DC Voltage last falls to 10% of its maximum amplitude.
The <source1> parameter is the AC Voltage and the <source2> parameter is
the DC Voltage. These sources can also be specified by the
:MEASure:SOURce command.
Query Syntax
:MEASure:OFFTime? [<source1>][,<source2>]
The :MEASure:OFFTime query returns the measured turn off time.
Return Format
<return_value><NL>
<return_value> ::= the time in seconds in NR3 format
See Also • ":MEASure:SOURce" on page 444
• ":POWer:ONOFf:TEST" on page 556
• ":POWer:ONOFf:APPLy" on page 553