:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 649
:SBUS<n>:A429:TRIGger:RANGe
(see page 1138)
Command Syntax
:SBUS<n>:A429:TRIGger:RANGe <min>,<max>
<min> ::= 8-bit integer in decimal, <hex>, <octal>, or <string>
from 0-255
<max> ::= 8-bit integer in decimal, <hex>, <octal>, or <string>
from 0-255
<hex> ::= #Hnn where n ::= {0,..,9 | A,..,F}
<octal> ::= #Qnnn where n ::= {0,..,7}
<string> ::= "0xnn" where n::= {0,..,9 | A,..,F}
The :SBUS<n>:A429:TRIGger:RANGe command defines a range of ARINC
429 label values. This range is used when the LRANge trigger type is
selected.
Query Syntax
:SBUS<n>:A429:TRIGger:RANGe?
The :SBUS<n>:A429:TRIGger:RANGe? query returns the current label values
in decimal format.
Return Format
<min>,<max><NL> in decimal format
Errors • "- 241, Hardware missing" on page 1099
See Also • "Introduction to :TRIGger Commands" on page 877
• ":SBUS<n>:A429:TRIGger:TYPE" on page 650

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals