270 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
10 :CHANnel<n> Commands
:CHANnel<n>:OFFSet
(see page 1138)
Command Syntax
:CHANnel<n>:OFFSet <offset> [<suffix>]
<offset> ::= Vertical offset value in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:OFFSet command sets the value that is represented at
center screen for the selected channel. The range of legal values varies
with the value set by the :CHANnel<n>:RANGe and :CHANnel<n>:SCALe
commands. If you set the offset to a value outside of the legal range, the
offset value is automatically set to the nearest legal value. Legal values are
affected by the probe attenuation setting.
Query Syntax
:CHANnel<n>:OFFSet?
The :CHANnel<n>:OFFSet? query returns the current offset value for the
selected channel.
Return Format
<offset><NL>
<offset> ::= Vertical offset value in NR3 format
See Also • "Introduction to :CHANnel<n> Commands" on page 262
• ":CHANnel<n>:RANGe" on page 277
• ":CHANnel<n>:SCALe" on page 278
• ":CHANnel<n>:PROBe" on page 271