790 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :SEARch Commands
:SEARch:GLITch:RANGe
(see page 1138)
Command Syntax
:SEARch:GLITch:RANGe <less_than_time>[suffix],
<greater_than_time>[suffix]
<less_than_time> ::= (15 ns - 10 seconds) in NR3 format
<greater_than_time> ::= (10 ns - 9.99 seconds) in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :SEARch:GLITch:RANGe command sets the pulse width duration for
the selected :SEARch:GLITch:SOURce. You can enter the parameters in any
order — the smaller value becomes the <greater_than_time> and the larger
value becomes the <less_than_time>.
Query Syntax
:SEARch:GLITch:RANGe?
The :SEARch:GLITch:RANGe? query returns the pulse width duration time
for :SEARch:GLITch:SOURce.
Return Format
<less_than_time>,<greater_than_time><NL>
See Also • Chapter 28, “:SEARch Commands,” starting on page 777
• ":SEARch:GLITch:SOURce" on page 791
• ":SEARch:GLITch:QUALifier" on page 789
• ":SEARch:MODE" on page 780