506 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
22 :MTESt Commands
:MTESt:ENABle
(see page 1138)
Command Syntax
:MTESt:ENABle <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:ENABle command enables or disables the mask test features.
• ON — Enables the mask test features.
• OFF — Disables the mask test features.
Query Syntax
:MTESt:ENABle?
The :MTESt:ENABle? query returns the current state of mask test features.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also • "Introduction to :MTESt Commands" on page 491