EasyManuals Logo
Home>Agilent Technologies>Test Equipment>InfiniiVision 3000 X-Series

Agilent Technologies InfiniiVision 3000 X-Series Programmer's Guide

Agilent Technologies InfiniiVision 3000 X-Series
1282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #442 background imageLoading...
Page #442 background image
442 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
20 :MEASure Commands
:MEASure:SDEViation
(see page 1138)
Command Syntax
:MEASure:SDEViation [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1-2 or 1-4 (# of analog channels) in NR1 format
<r> ::= 1-2 in NR1 format
The :MEASure:SDEViation command installs a screen measurement and
starts std deviation measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax
:MEASure:SDEViation? [<source>]
The :MEASure:SDEViation? query measures and outputs the std deviation
of the selected waveform. The oscilloscope computes the std deviation on
all displayed data points.
Return Format
<value><NL>
<value> ::= calculated std deviation value in NR3 format
See Also • "Introduction to :MEASure Commands" on page 406
• ":MEASure:VRMS" on page 463
• ":MEASure:SOURce" on page 444
NOTE
This ":MEASure:VRMS DISPlay, AC" command is the preferred syntax for making standard
deviation measurements.
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals