:SBUS<n> Commands 27
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 671
:SBUS<n>:FLEXray:AUTosetup
(see page 1138)
Command Syntax
:SBUS<n>:FLEXray:AUTosetup
The :SBUS<n>:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
• Sets the selected source channel's impedance to 50 Ohms.
• Sets the selected source channel' s probe attenuation to 10:1.
• Sets the trigger level (on the selected source channel) to - 300 mV.
• Turns on trigger Noise Reject.
• Turns on Serial Decode.
• Sets the trigger to the specified serial bus (n of SBUS<n>).
See Also • "Introduction to :TRIGger Commands" on page 877
• ":SBUS<n>:FLEXray:TRIGger" on page 679
• ":SBUS<n>:FLEXray:BAUDrate" on page 672
• ":TRIGger[:EDGE]:LEVel" on page 903
• ":SBUS<n>:FLEXray:SOURce" on page 678

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals