1022 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
33 :WGEN Commands
:WGEN:MODulation:STATe
(see page 1138)
Command Syntax
:WGEN:MODulation:STATe <setting>
<setting> ::= {{OFF | 0} | {ON | 1}}
The :WGEN:MODulation:STATe command enables or disables modulated
waveform generator output.
You can enable modulation for all waveform generator function types
except pulse, DC, and noise.
Query Syntax
:WGEN:MODulation:STATe?
The :WGEN:MODulation:STATe? query returns whether the modulated
waveform generator output is enabled of disabled.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also • ":WGEN:MODulation:AM:DEPTh" on page 1013
• ":WGEN:MODulation:AM:FREQuency" on page 1014
• ":WGEN:MODulation:FM:DEViation" on page 1015
• ":WGEN:MODulation:FM:FREQuency" on page 1016
• ":WGEN:MODulation:FSKey:FREQuency" on page 1017
• ":WGEN:MODulation:FSKey:RATE" on page 1018
• ":WGEN:MODulation:FUNCtion" on page 1019
• ":WGEN:MODulation:FUNCtion:RAMP:SYMMetry" on page 1020
• ":WGEN:MODulation:TYPE" on page 1023