:WGEN Commands 33
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 1021
:WGEN:MODulation:NOISe
(see page 1138)
Command Syntax
:WGEN:MODulation:NOISe <percent>
<percent> ::= 0 to 100
The :WGEN:MODulation:NOISe command adds noise to the currently
selected signal. The sum of the amplitude between the original signal and
injected noise is limited to the regular amplitude limit (for example, 5 Vpp
in 1 MOhm), so the range for <percent> varies according to current
amplitude.
Note that adding noise affects edge triggering on the waveform generator
source as well as the waveform generator sync pulse output signal (which
can be sent to TRIG OUT). This is because the trigger comparator is
located after the noise source.
Query Syntax
:WGEN:MODulation:NOISe?
The :WGEN:MODulation:NOISe query returns the percent of added noise.
Return Format
<percent><NL>
<percent> ::= 0 to 100
See Also • ":WGEN:FUNCtion" on page 1007

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals