:FUNCtion Commands 16
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 331
:FUNCtion:BUS:YORigin
(see page 1138)
Command Syntax
:FUNCtion:BUS:YORigin <value>
<value> ::= value at bus code = 0, in NR3 format
The :FUNCtion:BUS:YORigin command specifies the value associated with
Chart Logic Bus data equal to zero.
This command is available with the DSOX3ADVMATH advanced math
license.
Query Syntax
:FUNCtion:BUS:YORigin?
The :FUNCtion:BUS:YORigin query returns the value for associated with
data equal to zero.
Return Format
<value><NL>
<value> ::= value at bus code = 0, in NR3 format
See Also • ":FUNCtion:OPERation" on page 347