Obsolete and Discontinued Commands 35
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 1077
:MEASure:VDELta
(see page 1138)
Query Syntax
:MEASure:VDELta?
The :MEASure:VDELta? query returns the voltage difference between
vertical marker 1 (Y1 cursor) and vertical marker 2 (Y2 cursor). No
measurement is made when the :MEASure:VDELta? query is received by
the oscilloscope. The delta value that is returned is the current value. This
is the same value as the front- panel cursors delta Y value.
VDELta = value at marker 2 - value at marker 1
Return Format
<value><NL>
<value> ::= delta V value in NR1 format
See Also • "Introduction to :MARKer Commands" on page 380
• "Introduction to :MEASure Commands" on page 406
• ":MARKer:Y1Position" on page 389
• ":MARKer:Y2Position" on page 390
• ":MARKer:YDELta" on page 391
• ":MEASure:TDELta" on page 1068
• ":MEASure:TSTArt" on page 1072
NOTE
The :MEASure:VDELta command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:YDELta command (see page 391) instead.

Table of Contents

Other manuals for Agilent Technologies InfiniiVision 3000 X-Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 X-Series and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 X-Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000 X-Series
CategoryTest Equipment
LanguageEnglish

Related product manuals