:FUNCtion Commands 16
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 329
:FUNCtion:BUS:SLOPe
(see page 1138)
Command Syntax
:FUNCtion:BUS:SLOPe <slope>
<slope> ::= {NEGative | POSitive | EITHer}
The :FUNCtion:BUS:SLOPe command specifies the clock signal edge for the
Chart Logic Bus State operation.
This command is available with the DSOX3ADVMATH advanced math
license.
Query Syntax
:FUNCtion:BUS:SLOPe?
The :FUNCtion:BUS:SLOPe query returns the clock edge setting.
Return Format
<slope><NL>
<slope> ::= {NEGative | POSitive | EITHer}