6 Transesophageal Transducer
Instructions for Use 6 - 23
Temperature Sensor Failure (V5Ms)
WARNING: Disconnect the transducer if the system displays the message indicating a
temperature monitor failure. Remove the transducer from the patient. Failure to disconnect and
remove a transesophageal transducer with faulty temperature sensor can burn the patient's
esophagus. Contact your Siemens service representative.
Follow the instructions included in the on-screen message if the system detects a temperature
monitor failure:
Temperature Monitor Failure
Temperature monitor failure.
Disconnect transducer.
The system detects a
temperature sensor
failure.
1. Click Close to remove the message.
2. Disconnect the transducer.
3. Remove the transducer from the patient.
4. Contact your Siemens service
representative.