2 Safety and Care
Instructions for Use 2 - 17
Adjusting the Transmit Power during Contrast Agent Imaging
During contrast agent imaging, you can also increase or decrease the transmit power. The
transmit power range is from 0.10% to 100% for contrast agent imaging.
● Rotate the [Power] control located below the touch screen clockwise to increase transmit
power. Rotate the [Power] control counterclockwise to decrease transmit power.
○ For systems without a touch screen, rotate the [Power] LED panel control clockwise to
increase transmit power. Rotate the [Power] control counterclockwise to decrease transmit
power.
Transmit Power Display
The transmit power range is from 0% to 100%. Selecting 100%, in combination with other
ultrasound system controls or functions, generates the maximum acoustic intensity and
mechanical index for each transducer, where:
I
SPTA.3
: ≤ 720 mW/cm
2
and MI ≤ 1.9