2 Safety and Care
Instructions for Use 2 - 19
Transducer Surface Temperature Limits
The following table provides the maximum surface temperature of the transducers compatible
with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Maximum Temperature
Transducer
TMM
Still Air
4C1
≤
38.6°C
≤
35.9°C
6C1 HD
≤ 41.0°C ≤ 38.0°C
6C2
≤ 40.1°C ≤ 40.0°C
8C3 HD
≤ 41.0°C ≤ 42.5°C
9L4
≤ 39.4°C ≤ 38.1°C
12L4
≤ 42.5°C ≤ 47.0°C
14L5
≤ 41.7°C ≤ 41.1°C
14L5 SP
≤
40.7°C
≤
40.5°C
18L6 HD
≤ 42.9°C ≤ 43.0°C
4P1
≤ 40.9°C ≤ 42.6°C
4V1
≤ 40.3°C ≤ 40.8°C
4V1c
≤ 42.3°C ≤ 46.0°C
8V3
≤
41.4°C
≤
41.0°C
10V4
≤ 37.7°C ≤ 36.5°C
V5Ms
≤ 36.4°C ≤ 39.4°C
V7M
≤ 42.6°C ≤ 40.3°C
EV-8C4
≤ 38.9°C ≤ 39.6°C
EC9-4
≤
36.0°C
≤
36.7°C
MC9-4
≤ 42.5°C ≤ 45.3°C
7CF1
≤ 38.6°C ≤ 45.0°C
7CF2
≤ 36.7°C ≤ 37.2°C
9EVF4
≤ 38.4°C ≤ 38.3°C
CW2
≤ 36.9°C ≤ 37.7°C
CW5
≤ 35.3°C ≤ 37.7°C
AcuNav 8F
≤
41.1°C
NA
AcuNav 10F
≤ 39.6°C
NA
TMM = Tissue Mimicking Material
NA = not applicable (not tested)