6 Transesophageal Transducer
Instructions for Use 6 - 37
V5Ms Transducer Technical Description
Transducer Type: Phased Array, multiplane
Elements: 64
0° to 180°
Distal Tip Articulation: 120° anterior, 90° posterior
45° right, 45° left
Maximum insertion length:
1080 mm
15 mm
2D-mode Transmit
4.0 MHz, 5.0 MHz, 6.0 MHz, 7.0 MHz
Color mode Transmit
Frequencies, including
3.5 MHz, 3.75 MHz, 4.0 MHz
Doppler Transmit
Frequencies, including PW,
SCW, Aux CW,
Pulsed Wave DTI: 3.5 MHz, 3.75 MHz, 4.0 MHz
Imaging Modes:
2D-mode, PW Doppler, Steerable CW Doppler, Color mode, M-mode, Color DTI,
Pulsed Wave DTI, Anatomical M-mode
Maximum Image
Field of View: 90°
Sector
Operating Environment
Temperature:
Relative Humidity:
10°C to +40°C
10% to 80%
Storage Environment
Temperature:
Relative Humidity:
-10°C to +50°C
10% to 95%
Degree of protection
against electrical shock:
V5Ms transducers with this label are certified as type BF
defibrillator proof per IEC 60601-1 Standard for Safety of
Medical Equipment
V5Ms transducers with this label are certified as type BF (not
defibrillator proof) per IEC 60601-1 Standard for Safety of
Medical Equipment
Degree of protection
against ingress of fluid: IPX8