Appendix A Technical Description
A - 40 Instructions for Use
Environmental Requirements
EMC Note: Operating the ultrasound imaging system in close proximity to sources of strong
electromagnetic fields, such as radio transmitter stations or similar installations may lead to
interference visible on the monitor screen. However, the device has been designed and tested to
withstand such interference and will not be permanently damaged.
During Operation
Ambient (During Storage or
Transportation)
10% to 80%, non-condensing
10% to 95%, non-condensing
+10°C to +40°C* -20°C to +60°C
Array Transducers
+10°C to +40°C -10°C to +50°C
+20°C to +40°C -5°C to +50°C
Pencil CW Transducers
CW2, CW5
+10°C to +40°C* -40°C to +60°C
Maximum Altitude:
Operation up to 3,050 meters
5,050 meters (18,000 feet)
*For systems with documentation devices installed.
Protective Measures
Explosion protection:
This product is not designed for operation in areas subject to explosion
hazards.
Maximum Physical Dimensions
Width:
623 mm
(For systems with a
touch screen)
1,630 mm
1,330 mm (minimum)
(For systems without
a touch screen)
1,540 mm
1,275 mm (minimum)
Weight: Not to exceed 170 kg (375 lbs)
The weight of the system is based on the system unpacked and ready
for connecting transducers and using in a patient examination.
The weight includes documentation devices, such as printers, installed
on the system.