Commands listed in alphabetical order
Arguments
1 or ON turns on t
he AFG additive noise state.
0 or OFF turns it off.
Examples
AFG:NOISEADD:STATE ON turns on the additive noise state.
AFG:NOISEADD:STATE?mig
ht return
:AFG:NOISEADD:STATE 0 indicating
the n oise additive state is off.
AFG:OFFSet
Sets (or queries) the AFG offset value, in volts.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:OFFSet <NR3>
AFG:OFFSet?
Arguments
Floating point number that represents the AFG offset, in volts.
Examples
AFG:OFFSET 1.0 sets the AFG o ffset to 1.0 volts .
AFG:OFFSET? might return :AFG:OFFSET 0.0E+0 indicating there is no offset.
AFG:OUTPut:LOAd:IMPEDance
Sets (or queries) the AFG output load impedance.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:OUTPut:LOAd:IMPEDance {FIFty|HIGHZ }
AFG:OUTPut:LOAd:IMPEDance ?
Arguments
FIFty sets the output load impedance to 50 Ohms.
HIGHZ sets the output load impedance to the high-impedance state.
2-104 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO58

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO58 and is the answer not in the manual?

Tektronix MSO58 Specifications

General IconGeneral
BrandTektronix
ModelMSO58
CategoryTest Equipment
LanguageEnglish

Related product manuals