Commands listed in alphabetical order
Syntax
POWer:POWer<x
>:PERIOD:INPUTSOurce
{CH<x> | MATH<x> | REF<x>}
POWer:POWer<x>:PERIOD:INPU TSOurce?
Argu
ments
CH<x> = A channel specifier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:PERIOD:INPUTS Ource CH6 sets the input source for period
measurement as channel 6.
POWer:POWer<x>:POWERQUALITY:CCYCles
This command sets or queries the calculate cycles over full cycles settings for the
specified power quality measurement badge. The power measurement badge
is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:POWERQUALIT Y:CCYCles {ON|OFF1|0}
POWer:POWer<x>:POWERQUALIT Y:CCYCles?
Arguments
1|ONselects the calculate cycles over full cycles.
0 | OFF unselects the calculate cycles over full cycles.
Examples
POWer:POWer1:POWERQUALITY: CCYCles 1 selects the calculate cycles over
full cycles for the power measurement badge 1.
POWer:POWer<x>:POWERQUALITY:FREFerence
This command sets or queries the frequency reference type for power quality
measurement in the specified power measurement badge. The power measurement
badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-549